| EXHALYZER V - veterinarian spirometry |
| |
| Measurement principle |
Flow: ultrasonic transit time
|
|
| Flow related parameters |
|
| Time resolution |
10 ns |
| Sampling frequency |
up to 200 Hz |
|
Measurement range
|
100 l/s
|
| Detection limit |
6 ml/s |
| Dead space |
300 ml
|
|
| FRC measurement |
|
| Applications |
Cont. flow system |
|
Required gases
|
Cont. flow system: 25% Helium, 21% Oxygen balaced Nitrogen or 5% SF6, 21% Oxygen balanced Nitrogen
|
| Operating specifications |
|
| Temperature range |
10 to 40 °C |
|
Permitted humidity range
|
5 to 95% RH (non condensing) |
| Atmospheric pressure |
700 to 1060 hPa |
| Computer connection |
USB or RS 232 |
| Weight: |
Flow head: 1.2 kg Instrument: 5 kg |
|
Power uptake
|
400 VA max. |
| Mains voltage |
100 to 240 V, 50/60 Hz |
|
Dimensions
(Width x Height x Depth)
|
500 mm x 135 mm x 400 mm 19 x 5 ¼ x 17 inch
|
|
| PC System Requirements |
Pentium 3 type processor or higher, Microsoft Windows XP and Internet Explorer (V4.0 or higher), 16 MByte of RAM, 10MB of free disk space, USB port, XGA or better graphic adapter and RS232 interface
|
|
|