EXHALYZER V - veterinarian spirometry
 
Measurement principle
Flow: ultrasonic transit time

Flow related parameters  
Time resolution 10 ns
Sampling frequency up to 200 Hz
Measurement range
100 l/s
Detection limit 6 ml/s
Dead space
300 ml

FRC measurement  
Applications Cont. flow system
Required gases
Cont. flow system: 25% Helium, 21% Oxygen balaced Nitrogen or 5% SF6, 21% Oxygen balanced Nitrogen
Operating specifications  
Temperature range 10 to 40 °C
Permitted humidity range
5 to 95% RH (non condensing)
Atmospheric pressure 700 to 1060 hPa
Computer connection USB or RS 232
Weight: Flow head: 1.2 kg
Instrument: 5 kg
Power uptake
400 VA max.
Mains voltage 100 to 240 V, 50/60 Hz
Dimensions
(Width x Height x Depth)
500 mm x 135 mm x 400 mm
19 x 5 ¼ x 17 inch

PC System Requirements

Pentium 3 type processor or higher, Microsoft Windows XP and Internet Explorer (V4.0 or higher), 16 MByte of RAM, 10MB of free disk space, USB port, XGA or better graphic adapter and RS232 interface

 
ECO MEDICS reserves the right to change these specifications without notice