ANALYZER CLD 88sp - Nitric oxide analyzer with integrated ultrasonic flow measurement
 
Measurement principle

Nitric oxide: chemiluminescence (reference method)
Flow: ultrasonic transit time


NO related parameters
Measurement range 0.1 ppb to 5000 ppb
Detection limit < 0.06 ppb, sample tube type 3
< 0.1 ppb, sample tube type 1

Lag time
(software compensated)

< 0.9 s, sample tube type 3
< 2.7 s, sample tube type 1
Rise time (0-90%) < 0.1 sec
Fall time (0-90%) < 0.1 sec
Internal sampling rate 40 Hz
Data output rate 15 Hz
Linearity < 1% of full scale
Zero-point drift < 0.5 ppb / 6 h
Interference
» 3% of meas. value per vol.-% H2O
» 1% of meas. value per vol.-% CO2

Flow related parameters  
Time resolution 10 ns
Sampling frequency up to 200 Hz
Measurement range
up to 15 l/s
Detection limit 0.6 ml/s
Dead space

Set 1: 2 ml
Set 2: 10 ml


Operating specifications  
Temperature range 10 to 40°C
Permitted humidity range
5 to 95% RH (non condensing)
Sample flow rate 110 ml/min ±10 %, sample tube type1
330 ml/min ±10 %, sample tube type3
Cal. gas concentration 0.1 to 4 ppm NO in N2 (1 ppm recommended)
Cal. gas flow rate approx. 600 ml/min
Inlet supply 200 to 800 kPa (2 to 8 bar), 1/8’’ Swagelock
Power uptake
400 VA max
Mains voltage 100 to 240 V, 50/60 Hz
Dimensions
(Width x Height x Depth)

500 mm x 135 mm x 540 mm
19 x 5 ¼ x 17 inch

Weight 24 kg

PC System Requirements

Pentium 3 type processor or higher, compliant to the Council Directive 93/42 EEC concerning medical devices and the European Safety Standard EN 60601-1
Microsoft Windows XP, VISTA or 7, 16 MByte of RAM, 10MB of free disk space, USB port, XGA or better graphic adapter, RS232 interface

CE MDD 0483  
   
ECO MEDICS reserves the right to change these specifications without notice